Micro electronic intrinsic safe gage pressure transmitters with retasking measurement range MIDA-SA-13-Ex-CN
Applications industrial
instrumentation & process control systems, including atomic power stations
Measured medium fluids and gases, compatible with titanium alloys and stainless
steel Standard ranges, MPa 0-0,01; 0-0,016; 0-0,025; 0-0,04; 0-0,06; 0-0,1;
0-0,16; 0-0,25; 0-0,4; 0-0,6; 0-1; 0-1,6;0-2,5; 0-4; 0-6; 0-10; 0-16; 0-25;
0-40; 0-60; 0-100; 0-160. Overpressure 2x for ranges up to 10 MPa; 1,5x for
higher ranges Burst pressure, no less than 4x for ranges up to 10 MPa; 2x for
higher ranges. Overall accuracy in baseline measurement range ±% 0,25. Overall accuracy in retasking measurement range ±% 0,5. Operating temperature range, ° C -40 ... +80. Output 4-20 mA (2-wire).
Supply voltage, V d.c.. from intrinsic safe supply unit Intrinsic safe supply
units MIDA-PSSB-102-Ex, other compatible safe barrier systems Intrinsic safety
spark-protected electrical circuit; OExiaIICT4 . Ingress protection IP65.
Electrical connection Plastic sealing gland straight (S) and angle (A); metalic
sealing gland straight (SM) and angle (AM); plug/socket (P). Pressure port
M20x1,5; M12x1,5. Weight, no more than kg 0,25.
|